Abstract

This paper presents the design of a multi-axis capacitive tactile force sensor with a fully decoupled output response for input normal and shear forces. A patterned elastomer is used as a dielectric layer between capacitive electrodes of the sensor that allows to achieve relatively higher sensitivity. The sensor is fabricated utilizing a low-cost rapid prototyping technique and is characterized for normal and shear forces in the range of 0 ~ 10 N and 0 ~ 3.1 N respectively. The achieved force sensitivity for the normal axis is 2.03%/N and for shear axes is 1.67%/N. The difference between the estimated force from the sensor and actual force applied is negligible, which demonstrates the accuracy of the sensor. The reliability of the sensor is analysed by performing hysteresis and repeatability tests. The hysteresis error is found to be 4.94% and 4.69% for normal and shear forces respectively. The repeatability error of the sensor is less than 5%, which shows the stability of the sensor. The high sensitivity, linear output response, high force measurement range, reliability and low cost make the proposed tactile sensor suitable for the force feedback in the robotic surgical systems.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.