Abstract
In the development of new scintillators for X-ray imaging, a high-resolution and highly efficient system is required to evaluate the performance of the scintillator plates. For this purpose, we developed a high-resolution X-ray microscope system. The developed compact X-ray microscope system is based on a magnifying unit and a cooled charge-coupled device (CCD) camera, combined with a small industrial X-ray irradiation system. Using this system, we carried out imaging of three scintillator plates and evaluated their spatial resolution. Each scintillator plates was set in front of the lens of the objective, X-rays were irradiated to the scintillator plates, and transmission images of masks were acquired. The measured spatial resolution of the scintillator plates varied from 16 μm to 30 μm, depending on the type of scintillator plate. The focus size of the X-ray tube had an almost negligible effect on the spatial resolution of the images for the evaluated scintillator plates.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.