Abstract

Front end electronics (detector to ADC) have been developed to operate over an input count rate range of 105 to 106 counts per second with less than ± 1% gain shift and minimal resolution degradation. The system was developed to satisfy requirements for a nuclear hardened gamma densitometer utilizing multichannel analyzer pulse height analysis techniques for reactor background subtraction. The circuitry developed includes a fast current mode preamplifier with first order count rate dependent gain compensation, a baseline restorer, a single channel analyzer with pile-up rejection and live time measuring circuits, and a fast linear gate. The single channel analyzer is used for energy region of interest selection so that only those pulses within the energy region of interest are presented to the analog to digital converter.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.