Abstract

Abstract Breast cancer is the most commonly diagnosed cancer in women. A strong treatment candidate is high-intensity focused ultrasound (HIFU), a non-invasive therapeutic method that has already demonstrated its promise. To improve the precision and lower the cost of HIFU treatment, our group has developed an ultrasound (US)-guided, five-degree-of-freedom (DOF), robot-assisted HIFU system. We constructed a fully functional prototype enabling easy three-dimensional (3D) US image reconstruction, target segmentation, treatment path generation, and automatic HIFU irradiation. The position was calibrated using a wire phantom and the coagulated area was assessed on heterogeneous tissue phantoms. Under the US guidance, the centroids of the HIFU-ablated area deviated by less than 2 mm from the planned treatment region. The overshoot around the planned region was well below the tolerance of clinical usage. Our system is considered to be sufficiently accurate for breast cancer treatment.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.