Abstract

Changes in limb volume and shape among transtibial amputees affects socket fit and comfort. The ability to accurately measure residual limb volume and shape and relate it to comfort could contribute to advances in socket design and overall care. This work designed and validated a novel 3D laser scanner that measures the volume and shape of residual limbs. The system was designed to provide accurate and repeatable scans, minimize scan duration, and account for limb motion during scans. The scanner was first validated using a cylindrical body with a known shape. Mean volumetric errors of 0.17% were found under static conditions, corresponding to a radial spatial resolution of 0.1 mm. Limb scans were also performed on a transtibial amputee and yielded a standard deviation of 8.1 ml (0.7%) across five scans, and a 46 ml (4%) change in limb volume when the socket was doffed after 15 minutes of standing.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.