Abstract

During measuring the thermal contact resistance (TCR) of solids by using traditional steady-state test methods, it is observed that the measured results are unavoidable to be impacted by the direction of exerting heat flux between dissimilar and similar metals. Such a directional effect has also been found for the cases between two solids of identical material with the same surface properties. Uncertainty analysis shows that the directional effect between two solids may result from the additive errors. In order to improve the measurement precision, we propose a method by using harmonic mean value of TCR in two directions of exerting heat flux to diminish the effect. Meanwhile, a modified experimental apparatus has been designed and established to high-precisely measure TCR between two solids of the identical material. To verify the accuracy of the method, the TCRs of samples of 99.999% standard pure copper and Elkonite copper–tungsten alloy 30W3 are measured and discussed in detail. The results show that the present method has high precision and can be used in a relatively wide range of TCR measurement for identical solids and the characterization of thermal interface materials.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.