Abstract

An automatic apparatus for the continuous measurement of O2 consumption of small laboratory animals is described. By use of a high-sensitivity pressure transducer with associated circuitry together with a peristaltic O2 delivery system, the closed respirometer chamber is maintained at atmospheric pressure +/- 0.5 mmH2O. O2 delivery is measured to within 0.25 ml by recording rotations of the peristaltic pump, following calibration by the withdrawal of a preset volume of air from the chamber. Static trials (with the chamber empty) indicate a high degree of reproducibility of data with the chamber pressure remaining at atmospheric pressure +/- 0.5 mmH2O as a result of the proportional, as opposed to fixed-volume, delivery of O2. Trials with mice and rats have likewise produced data with a high degree of reproducibility.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.