Abstract

The design, calibration, and performance of a high-power (1-1000 W) automatic network analyzer based on the six-port concept are described for the 10-100-MHz range. Calibration is performed with a length of transmission line as the only impedence standard needed. A 10-mW thermistor mount is the standard of power. Imprecision in measnring reflection coefficient Gamma is 0.0001 in magnitude and 0.005/|Gamma| degrees in phase. Corresponding estimated systematic errors are 0.001 and 0.1/|Gamma| degrees. Imprecision in measuring power is 0.01 percent of range (20 W, 200 W, or 1000 W) with an estimated systematic error of 1.25 percent of reading.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.