Abstract
We report preliminary testing of "GeneTrack", an instrument designed for the specific application of multiplexed short tandem repeat (STR) DNA analysis. The system supports a glass microdevice with 16 lanes of 20 cm effective length and double-T cross injectors. A high-speed galvanometer-scanned four-color detector was specially designed to accommodate the high elution rates on the microdevice. All aspects of the system were carefully matched to practical crime lab requirements for rapid reproducible analysis of crime-scene DNA evidence in conjunction with the United States DNA database (CODIS). Statistically significant studies demonstrate that an absolute, three-sigma, peak accuracy of 0.4-0.9 base pair (bp) can be achieved for the CODIS 13-locus multiplex, utilizing a single channel per sample. Only 0.5 microL of PCR product is needed per lane, a significant reduction in the consumption of costly chemicals in comparison to commercial capillary machines. The instrument is also designed to address problems in temperature-dependent decalibration and environmental sensitivity, which are weaknesses of the commercial capillary machines for the forensics application.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.