A gradient high-performance liquid chromatography micromethod is described for the simultaneous quantitation of vigabatrin and gabapentin in human serum. Chromatography was performed using a 125- x 3-mm ID Hypersil BDS C-18 column with a 3-microm mini-bore, eluted with a gradient system comprised of phosphate buffer (pH 6.5)-acetonitrile-methanol-water at a flow rate of 0.45 ml/minute. The column eluent was monitored on a fluorescence detector using excitation and emission wavelengths of 340 and 440 nm, respectively. The lower limit of quantitation for vigabatrin and for gabapentin was 5 micromol/l, and the within-batch and between-batch coefficients of variation were <5%. No interference from commonly prescribed antiepileptic drugs (carbamazepine and its metabolite carbamazepine epoxide, oxcarbazepine and its metabolite 10-hydroxycarbazepine, ethosuximide, lamotrigine, phenobarbitone, phenytoin, primidone, and valproic acid) was observed; thus, the method can be used to monitor vigabatrin and gabapentin in patients on polytherapy antiepileptic drug regimens.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.