Abstract

Background: We investigated whether the heparin-coated AN69 ST hemodialysis (HD) filter induced less hypercoagulability during HD than a conventional polysulfone filter (F×8). Methods: In a crossover design, 11 patients were treated alternately with AN69 ST and F×8 filters (45 sessions). All filters were primed with unfractionated heparin (UFH) and unadsorbed UFH was removed by saline flushing. Half the conventional dalteparin dose was given as a bolus dose at the start of HD. Clotting was evaluated hourly in the venous air trap. Prothrombin fragments 1 and 2 (PF1 + 2), antithrombin (AT), β-TG and anti-FXa activity were repeatedly measured. Results: One patient treated with enalapril had two repeated adverse reactions to the AN69 ST filter and was excluded from the study. Use of the AN69 ST filter did not decrease the mean clot score or PF1 + 2, but decreased β-TG compared to the F×8 filter. Conclusion: The heparin-coated AN69 ST filter did not induce less coagulation when compared to the F×8 filter.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.