Abstract

Photoelectron spectrometers usually allow detection of either spin-resolved energy-distribution curves (EDCs) at single emission angle, or 2D angle- vs.-energy images without spin-resolution. We have combined the two detection schemes into one spectrometer system which permits simultaneous detection of a 1D spin-resolved EDC and a 2D angular map. A state-of-the-art hemispherical analyzer is used as an energy filter. Its original scintillator detector has been replaced by a delay-line-detector (DLD), and part of the electron beam is allowed to pass through to reach the spin-polarized low energy electron diffraction (SPLEED) spin-detector mounted subsequently. The electron-optics between DLD and SPLEED contains a 90 ° deflector to feature simultaneous detection of in-plane and out-of-plane spin components. These electron-optics have been optimized for high transmission to reduce acquisition times in the spin-resolved mode.