Abstract
Many of the newer models of mechanical ventilators include the pressure support (PS) mode of ventilation as a standard feature. The mechanism for the termination of a pressure support breath varies considerably among the various ventilators. The potential for the sudden inadvertent application of a high, unremitting pressure support breath or continuous positive airway pressure (CPAP) clearly exists with some ventilators and is illustrated in two cases. The potentially adverse consequences of high levels of CPAP are predictable and have been well described. Standard ventilator monitoring systems which do not include mean airway pressure monitoring may not detect this problem before significant adverse consequences have occurred. Even when detected, this unfamiliar complication may be difficult to recognize by the uninitiated. Physicians utilizing this mode of ventilation must be familiar with the terminating mechanism of their particular ventilator and choose appropriate monitoring.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.