Abstract

A case of cardiac arrest is presented which was caused by improper connection of a modified Mapleson D circuit (Bain breathing circuit). Excessive enternal deadspace was created by interchanging the gas inflow line and the attachment for an airway pressure manometer. This resulted in a marked respiratory acidosis, clinically undetected until the concomitant hypoxia produced a severe cardiac depression and arrest. The sequence of events was reproduced in a dog under comparable anaesthetic conditions. In order to avoid this error it is recommended to permanently fuse the connecting piece placed in the circuit with the tubing leading to the airway pressure manometer of the respirator.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.