Abstract

A special-purpose hardware architecture is proposed to implement the expectation-maximization algorithm to compute, in clinically useful times, the maximum-likelihood estimate of a radionuclide distribution for a positron-emission tomogram from time-of-flight measurements. Two-dimensional convolutions required for forming the estimate are converted into a series of one-dimensional convolutions that can be evaluated in parallel. Each one-dimensional convolution is evaluated using a number-theoretic transform. All numerical calculations are performed using finite-field arithmetic. To avoid the use of large finite fields and to increase parallelism, each convolution is performed by a series of convolutions with small digits in a Galois field. The hardware architecture allows a two-dimensional convolution of dimension NxN to be formed with O(N) operations, permitting one iteration of the expectation-maximization algorithm to be formed in approximately 100 ms when N=128 and the number of view angles is 32.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.