Abstract

Optical coherence tomography angiography (OCTA) has successfully demonstrated its viability for clinical applications in dermatology. Due to the high optical scattering property of skin, extracting high-quality OCTA images from skin tissues requires at least six-repeated scans. While the motion artifacts from the patient and the free hand-held probe can lead to a low-quality OCTA image. Our deep-learning-based scan pipeline enables fast and high-quality OCTA imaging with 0.3-s data acquisition. We utilize a fast scanning protocol with a 60 μm/pixel spatial interval rate and introduce angiography-reconstruction-transformer (ART) for 4× super-resolution of low transverse resolution OCTA images. The ART outperforms state-of-the-art networks in OCTA image super-resolution and provides a lighter network size. ART can restore microvessels while reducing the processing time by 85%, and maintaining improvements in structural similarity and peak-signal-to-noise ratio. This study represents that ART can achieve fast and flexible skin OCTA imaging while maintaining image quality.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.