A fibre-coded linear scintillator detector with a half-millimetre spatial resolution was developed with wavelength-shifting fibre read-out for neutron detection. The head of the detector comprised a ZnS/6LiF scintillator placed on the ribbon of wavelength-shifting (WLS) fibres. Both ends of the fibres were coded to reduce the number of read-out photomultipliers and associated electronic circuitry. The prototype detector in which the fibre ribbon was sandwiched between two ZnS/6LiF scintillators exhibited a spatial resolution of 0.5mm, a detection efficiency of 35% for 1-Å neutrons, and a gamma sensitivity of less than 10−7. This type of detector would be useful where a submillimetre spatial resolution is required with large detector coverage such as in neutron reflectometers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.