Abstract

Spectrally encoded confocal microscopy (SECM) is a novel approach for obtaining high resolution, depth-sectioned images of microstructure within turbid samples. By encoding one spatial dimension in wavelength, imaging probes can be greatly simplified compared to standard scanning confocal microscopes, potentially enabling endoscopic implementation. The use of a diffraction grating for spectral encoding, however, skews the optical axis through the probe, thus complicating the design of narrow diameter instruments. In this Letter, we describe a novel use of a single-optical-axis element based on high index-of-refraction prisms and a transmission holographic grating for the design of narrow diameter SECM devices. Confocal images obtained with a 10.0 mm probe demonstrate a transverse resolution of 1.1 microm and a field of view of 650 microm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.