Abstract

We describe the design, construction, and operation of a simple glow discharge unit that can be used to make surfaces such as carbon-coated electron microscopy grids and glass coverslips hydrophilic. The use of a vacuum leak detector (Tesla coil) in place of a conventional high-voltage power supply and a small plastic desiccator for the vacuum chamber make the unit very inexpensive. Owing to the small volume of the chamber and the simplicity of the unit, the whole glow discharge process can be carried out in only 2 to 3 min, a time considerably shorter than that required for conventional vacuum evaporators. The hydrophilic surface improves adsorption of particles by several orders of magnitude in preparation for negative staining, freeze-drying, and other procedures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.