Porous gold nanosheets modified glassy carbon electrode (GCE) was facilely prepared by one-step electrodeposition, using N-methylimidazole as a growth-directing agent. The porous gold nanosheets modified GCE was characterized by scanning electron microscopy, transmission electron microscopy, and X-ray diffraction spectroscopy. The modified electrode displayed improved sensitivity for individual and simultaneous differential pulse voltammetric determination of dopamine (DA; at 180 mV) and acetaminophen (AC; at 450 mV vs. Ag/AgCl) even in the presence of ascorbic acid. The oxidation peak currents linearly increased with the concentrations of DA and AC in the ranges from 2.0 to 298.0 μM and 3.0 to 320.0 μM, respectively, and the detection limits are 0.28 μM for DA and 0.23 μM for AC. The relative standard deviations (n = 20) are 1.5 % for DA and 0.4 % for AC.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.