Abstract

It was revealed that nominal mass spectra extracted from the same NetCDF file using different gas chromatography/mass spectrometry (GC/MS) software products are not identical. This phenomenon is caused by differences in algorithms used for rounding floating-point m/z values to integers. It was found that all programs under consideration (AMDIS, ChemStation, ChromaTOF, MS Search, OpenChrom) use different procedures. It is necessary to know how fractional parts of accurate m/z values of ions are distributed to determine which algorithm yields more robust results. We estimated the respective distribution using two databases (PubChem and NIST). As a result, we came up with a procedure that minimizes the influence of random errors on rounding to integer m/z values. The procedure we suggest is to sum intensities of all floating-point m/z values in a bin [MZ - 0.38; MZ + 0.62] and assign MZ as a nominal m/z value, where MZ is an integer m/z value.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.