Abstract

The aim of this study was to validate a Geant4 Monte Carlo application for dosimetry measurements at low energies when using water and PMMA phantoms. To measure the water-equivalence in plastic materials, relative dosimetry can be performed by deploying the calculation of depth dose profiles and output factors. We simulated a quasi-uniform monochromatic X-ray beam, with dimension of 2x2 cm2, perpendicularly impinging on a phantom, with energy ranging from 30 to 140 keV. A comparison of the attenuation of the beam after passing through slabs of PMMA and/or water is presented and validated against experimental measurement performed at the European Synchrotron Radiation Facility (ESRF) in Grenoble, France. The use of PMMA as a substitute of water at low energies is not recommended for dosimetric studies as the depth dose profiles show discrepancies of at least 3%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.