AbstractThe design and features of a sealed‐off proportional scintillation gas‐filled detector (GSPC) are described. The use of a 50 µm Be window and an Xe‐based gas mixture provides high efficiency of registration in the range 1.5–25 keV. The energy resolution changes from 18% for Al Kα (1.49 keV) to 4.2% for Ag Kα (22.2 keV). When the count rate on the unit output is 5 × 104 s−1, the energy resolution at the Mn Kα line does not exceed 9% at an entrance window area near 3 cm2. The detector is operated at room temperature and does not require maintenance during operation. The service life of the detector is over 5 years and the radiation lifetime is practically unlimited. The detector is designed for use in EDXRF spectrometers instead of gas‐filled proportional counters. Copyright © 2005 John Wiley & Sons, Ltd.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.