Abstract

A method for the determination of polybrominated diphenyl ethers (PBDEs) in biota for routine analysis is described. The mass spectroscopic (MS) evaluation of 23 brominated diphenyl ethers, under electron ionization and electron capture negative ion conditions using magnetic sector and quadrupole mass spectrometers, showed that high-resolution mass spectrometry (HRMS) under electron ionization conditions was the most reliable technique, with high selectivity and adequate sensitivity. The instrument detection limit for this method ranged for individual congeners between 4.8 and 0.1 pg for 3-bromodiphenyl ether (BDE-2) and 2,3 ′,4,4 ′-tetrabromodiphenyl ether (BDE-66), respectively, and method detection limit for each homologue group ranged between 5 pg/g for salmon certified reference material (CRM) and 93 pg/g for lake trout CRM. The effectiveness of this method was evaluated by analyzing the occurrence of PBDEs in commercially available CRMs comprising Lake Ontario lake trout, Pacific herring, and sockeye salmon. The average coefficients of variation for the replicate analyses of PDBEs in several tissue samples were: 25% for lake trout, 36% for Pacific herring, and 34% for sockeye salmon. The average deviations in the inter-laboratory study were: 14% for lake trout, 15% for Pacific herring, and 37% for sockeye salmon. Results indicated that the described method, based on gas chromatography/high-resolution mass spectrometry, is reliable for determining PBDE concentrations in biological tissues.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call