Abstract

A gamma-spectrometric method using an intrinsic high resolution germanium detector has been developed for the determination of isotope ratios of plutonium from samples in solution form. The method is based on the assay of low energy gamma-rays of238Pu,239Pu,240Pu and241Pu and does not require the use of branching intensities or the knowledge of detection efficiencies for different gamma rays. Since low energy gamma-rays are used, the effect of241Am has also been studied. It is found that results are not affected up to 0.5 wt% of241Am in plutonium samples. An accuracy of 3% is achievable in the determination of240Pu/239Pu and241Pu/239Pu atom ratios as demonstrated by carrying out measurements on isotopic standards of plutonium.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.