Abstract

The sensor is based on silver and platinum electrodes with an intervening silver iodide disk as a solid electrolyte. The small disk (13 mm diameter) is easily made by pressing (7000 kg cm -2) silver powder, silver iodide and a platinum gauze in layers in a die. The detector cell containing the disk is thermostated at 120 ± 0.1°C. When sample gas at 30 ml min -1 impinges on the platinum cathode, the current flowing in the external circuit is linearly related to the concentration of ozone and/or nitrogen dioxide up to 0.5 ppm from the detection limit of 0.5 ppb for ozone or 1 ppb for nitrogen dioxide.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.