Abstract

Rabbit anti-thymocyte globulin (rATG) has been widely used to prevent graft-versus-host disease (GvHD) after allogeneic hematopoietic stem cell transplantation (allo-HSCT). The therapeutic window of rATG is narrow, and it may increase the risk of relapse, viral reactivation, delayed immune reconstitution and GvHD when overexposed or underexposed. Therefore, a reliable method for detecting the rATG concentration in human serum by flow cytometry was established and fully validated for therapeutic drug monitoring. In this method, Jurkat T cells were used to capture active rATG in human serum, and PE-labeled donkey anti-rabbit IgG was used as a secondary antibody. The method showed good specificity, selectivity and excellent linearity at concentration of 0.00300–20.0 AU/mL. The intra- and interday precision values were all within 20% at four concentration levels for the analyte. The stock solutions of rATG showed no significant degradation after storage at ambient temperature for 8 h and at − 80 °C for 481 days. No significant degradation of rATG in serum was observed at ambient temperature for 6 h, during six freezethaw cycles and at − 80 °C for at least 373 days. This method was fully validated and successfully applied to monitor active rATG concentration in serum of patients with haploid-identical hematopoietic stem cell transplantation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.