Abstract

In this present study, a fully integrated BioMEMS was developed using silicon technology to simultaneously detect varying cytokine biomarkers: interleukin-1 (IL-1), interleukin-10 (IL-10), and interleukin-6 (IL-6) using eight gold working microelectrodes (WE). The biomarkers are one of many antigens that are secreted in acute stages of inflammation after left ventricle assisted device (LVAD) implantation for patients suffering from heart failure (HF). The monoclonal antibodies (mAb): anti-human IL-1, IL-10, and IL-6 were immobilized onto gold microelectrodes through functionalization with carboxyl diazonium, respectively. Cyclic voltammetry (CV) was applied during the microelectrode functionalization process to characterize the gold microelectrode surface properties, while electrochemical impedance spectroscopy (EIS) was used to characterize the modified gold microelectrodes. The BioMEMS was highly sensitive towards the three cytokines in a range of 1 pg/mL to 15 pg/mL, which is the window where acute inflammations were observed.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call