Abstract
A colorimetric method is described for determining total serum iron and unsaturated iron binding capacity (UIBC) of serum using Technicon AutoAnalyzer II equipment. The determination of UIBC has been fully automated by means of a simple modification to the Sampler II module, whereby 2 double probes permit the simultaneous aspiration of a sample and of a measured volume of the iron saturating solution. No manifold changes are needed when changing from serum iron to UIBC.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.