Abstract

A fully-automated, on-line, real-time analyzer has been developed for preconcentration and analysis of haloacetic acids (HAAs). Preconcentration of HAAs is achieved by sample acidification and solid phase extraction onto a hydrophobic polymeric resin using sequential injection analysis (SIA). The HAAs preconcentrate is then analyzed using post-column reaction-ion chromatography (PCR-IC), which is selective for HAAs. Systematic optimization of SIA preconcentration parameters are described followed by detailed method detection limit (MDL), accuracy, precision, and linearity studies. MDL values for the individual HAA9 species range from 0.4 to 0.9μgL−1. Side-by-side comparison studies of HAAs analysis in 14 real-world drinking water samples from Alabama, Arkansas, Kentucky, Minnesota, Missouri, Mississippi, New York, Pennsylvania and Tennessee are presented that compare the optimized SIA-PCR-IC to USEPA Method 552.3. Trace levels of HAAs detected in select samples are reported, and the bias values calculated between the two methods are typically less than 5μgL−1 for eight of the nine individual HAAs.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.