Abstract

PurposeThe objective of this work was to outline and demonstrate a standardized framework for evaluating automatically propagated contour quality against expert contours. A 2-pronged approach is used to evaluate contour quality: a geometric evaluation to identify geometric and spatial discrepancies between propagated and expert contours, and a comprehensive dosimetric comparison to provide clinical context for the results. Methods and MaterialsThe standardized framework requires a primary image, with reference contours and a radiation therapy treatment plan, and a secondary image. Reference contours are automatically propagated onto the secondary image anatomy and compared with expert contours obtained in an interobserver study. The standardized framework outlines geometric and dosimetric evaluation methodologies for determining indistinguishability between propagated and expert contours in a cohort analysis. Propagated contours are geometrically compared with expert contours in terms of the Dice similarity coefficient and the mean distance to agreement. Statistical analysis is performed on the central tendency and variability of Dice similarity coefficient and mean distance to agreement values over the patient cohort. Dosimetric evaluation involves computing the mean and 95% confidence intervals for the differences in cumulative dose-volume histograms for propagated and expert contours. A case study in accelerated partial breast irradiation was shown to demonstrate the framework. ResultsThe standardized framework was applied to a case study of 24 patient data sets with 3 radiation oncologists providing the expert contours. Cohort analysis indicated that propagated contours were geometrically indistinguishable and dosimetrically distinguishable from expert contours. ConclusionsThe recommended framework standardizes the comparison of geometric and dosimetric parameters to demonstrate indistinguishability of propagated contours from expert contours. Adoption of this framework is vital for consistent and comprehensive validation of automatic contour propagation for use in large-scale cohort analyses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.