Abstract

We propose in the present work how the reference glass dosimeters can be introduced, which reflects the user irradiation condition. The reference glass dosimeters are used for correcting the reader fluctuation by reading it with sample glass dosimeters at the same time. Since they can be used without annealing after irradiation for long periods, one should consider both the fading effect and the natural background dose accumulation quantitatively. We construct an empirical but practical formalism of evaluating the absorbed dose on the glass dosimeter with the fading effect and the natural background dose accumulation considered.