Abstract
An intramedullary rod for the femur was designed and compared with rods commonly used. Its bending strength was increased 40 to 80 per cent and its torsional rigidity was increased 230 to 3,000 per cent compared with the other devices tested. In forty patients with traumatic and pathological fractures and osteotomies of the femur, excellent fixation was achieved and very early weight-bearing was possible. In all fresh fractures and non-infected non-unions, union occurred very promptly and bone-grafting was not required in most instances. The stregth and rigidity of the device made it exceptionally useful for non-unions, pathological fractures of the femur, and resection-fusions of the knee.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.