Abstract

The isoenzyme creatine kinase MB is very important for diagnosis of acute myocardial infarction (AMI). Some CK-MB immunoassays are sensitive, accurate and available for clinical application, but they are expensive and time-consuming procedures. Furthermore, conventional fluorescence immunochromatographic assays (FL-ICAs) have suffered from background fluorescence interference and low analytical sensitivity. A rapid and simple FL-ICA with Eu (III) chelate polystyrene microparticles was developed to determine CK-MB in 50uL serum samples using a portable test strip reader by measuring the fluorescence peak heights of the test line (HT) and the control line (HC) in 12min. The assay was reliable with a good correlation coefficient between HT/HC ratio and CK-MB concentration in samples. A linear range was 0.85-100.29ng/mL for CK-MB, and the LOD was 0.029ng/mL. The intra- and inter-assay coefficients of variation (CV) were both <10% and the average recoveries were from 90.17% -112.63% for CK-MB. The system performed well in interference experiments. Furthermore, a highly significant correlation (r=0.9794, P<0.001) between this method and the commercially available bioMérieux mini VIDAS system were attained for measuring 120 CK-MB samples. These results indicated that the Eu (III) chelate microparticles-based FL-ICA is simple, fast, highly sensitive, reliable, and reproducible for point-of-care testing of CK-MB concentrations in serum. Graphical Abstract ᅟ.