The determination of angiotensin-converting enzyme (ACE) activity represents a useful tool in the study of different health pathologies, such as hypertension. This protocol describes a fluorescent assay for measuring ACE activity in vitro with high precision and sensitivity. The method relies on the ability of ACE to hydrolyse the internally quenched fluorescent substrate o-aminobenzoylglycyl-p-nitro-L-phenylalanyl-L-proline. The generation of the fluorescent product o-aminobenzoylglycine can be continuously monitored, preferably using a microtiter-plate fluorometer, though the use of a conventional cuvette fluorometer would also be possible. The method has important advantages with respect to other assays, because it involves only a one-step reagent, is easy to carry out and allows the analysis of an elevated number of samples in shorter times. It can be completed in one and a half hours. In addition, the fact that all reagents are commercially available allows the rapid introduction of the assay into the laboratory.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.