Abstract
The need for a flow-through detector with high efficiency and vanishingly low background arises in HPLC (high-pressure liquid chromatography) analysis of PET (positron emission tomography) tracer metabolites, where the free drug input function must be separated from the total circulating activity in sequential blood samples. Subnanocurie activities of positron emitters in 100 mu l plasma samples are chemically resolved, and then assayed by a triple coincidence beta - gamma - gamma -detector. Multiple coincidence logic between a BC-400 plastic scintillator flow-cell, sandwiched between two large NaI(Tl) gamma detectors, reduces the background event rate to 10/sup -3/ to 10/sup -6/ counts per second, with an overall efficiency of 25% for /sup 18/F-labeled eluates. Careful attention is paid to the construction of the beta cell, maintaining the chemical resolution of the HPLC system while optically transporting the few beta scintillation photons to multiple photomultipliers without losing the solid angle to the surrounding mantle of gamma detectors.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.