Abstract

The platelets (PLTs) in PLT concentrates are counted with hematology analyzers, but varying results among different hematology analyzers are observed, making comparisons very difficult. Due to the absence of red blood cells in PLT concentrates, the International Council for Standardization in Hematology (ICSH) reference method was modified to be used for PLT concentrates and validated in an international comparative study. Five PLT samples were shipped to eight participating centers of the Biomedical Excellence for Safer Transfusion (BEST) Collaborative and counted on the same day. PLTs were stained with fluorescein isothiocyanate-labeled anti-CD41a in tubes (TruCount, BD Biosciences), measured on a flow cytometer, and analyzed with a uniform template. These samples were also counted on 15 hematology analyzers. The ICSH method and newly developed BEST method yielded PLT counting results with less than 1% difference (not significant). The intercenter coefficient of variation (CV) of the BEST method was on average 6.3% versus 7.6% on average for hematology analyzers. The CV of individual hematology analyzers was on average 0.9%, which was considerably lower than for the flow cytometers with a mean of 3.7%. The BEST flow cytometric method has a smaller intercenter CV and a smaller center-to-center deviation from the group mean compared to hematology analyzers. Conversely, individual hematology analyzers are more precise than the flow cytometric method. Thus, the flow cytometric method provides a calibration tool to allow comparisons between centers, but there is no need to replace routine counting with hematology analyzers.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call