Abstract

A flow cytometry method using the halide-specific fluorescent dye, 6-methoxy-N-(3-sulfopropyl) quinolinium (SPQ), has been developed to measure intracellular chloride concentration in single cells. Collisions with chloride quench the fluorescence of SPQ, making it possible to relate the measured fluorescence intensity to chloride concentration with a Stern-Volmer equation. To demonstrate the method, porcine lymphocytes were loaded in vitro, using a hypotonic method, with 5 mM SPQ. Fluorescence excitation was provided by a UV laser and the fluorescence emission intensity at 485 nm was recorded. Calibration was performed by using 7 microM nigericin (a K/H antiporter) and 10 microM tributyltin (a Cl/OH antiporter) to equilibrate the concentrations of intracellular and extracellular chloride. Calibration measurements were made for chloride concentrations between 0 mM and 140 mM. The calibration produced a Stern-Volmer quenching constant of 16.2 M(-1) which was used to relate measured cell fluorescence to intracellular chloride concentration. The intracellular chloride concentration for fresh porcine lymphocytes was determined to be 56.2 +/- 3.3 mM. Stable loading of cells with 5 mM SPQ was accomplished in 15 minutes, leakage of SPQ from the cells was minimal, and over 95% of the cells remained viable after loading.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.