A flatness monitor has been built to quickly and accurately check accelerator beam flatness and dose calibration. Consisting of a 7 × 7 ion chamber array, the unit operates in photon beams from 60Co energies to 25 MV and electron beams (scattered or scanned) from 6 MeV to 25 MeV.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.