Abstract

Prosthetic alignment, patient characteristics, and implant design are all factors in long-term survival of total knee arthroplasty (TKA), yet the level at which each of these factors contribute to implant loosening has not been fully described. Prior clinical and biomechanical studies have indicated tibial overload as a cause of early TKA revision. The purpose of this study was to determine the relationship between tibial component design and bone resection on tibial loading. Finite-element analysis (FEA) was performed after simulated implantation of metal backed (MB) and all-polyethylene (AP) TKA components in 5 and 15 mm of tibial resection into a validated intact tibia model. Proximal tibial strains significantly increased between 13% and 199% when implanted with AP components (p < 0.05). Strain significantly increased between 12% and 209% in the posterior tibial compartment with increased bone resection (p < 0.05). This study indicates elevated strains in AP implanted tibias across the entirety of the proximal tibial cortex, as well as a posterior shift in tibial loading in instances of increased resection depth. These results are consistent with trends observed in prior biomechanical studies and may associate the documented device history of tibial collapse in AP components with increased bone strain and overload beneath the prosthesis.