Abstract

Commercially available cardiac pacemakers and implantable cardioverters/defibrillators (ICDs) predominantly use an intracardiac-derived electrocardiogram (ECG) for the detection of arrhythmias. To achieve automatic control of the heart frequency in accordance with cardiovascular strain and improved detection of life-threatening arrhythmias, it is desirable to monitor the heart by an input signal correlated with the hemodynamic state. One possible approach to derive such a signal is to measure the inotropy (mechanical contraction strength of the heart muscle). For this purpose, an optoelectronic measurement system has been designed. The fundamental function of the system has been shown in earlier investigations using an isolated beating pig heart. In this paper the design of two algorithms for use in pacemakers and ICDs based on a fiber optic sensor signal is presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.