Abstract

This article discusses how to monitor the freezing depth during cryotherapy using a fiber optic array sensor. The sensor was used to measure the backscattered and transmitted light from frozen and unfrozen ex vivo porcine tissue and in vivo human skin tissue (finger). The technique exploited the variations in optical diffusion properties of the frozen and unfrozen tissues to determine the extent of freezing. Ex vivo and in vivo measurements yielded comparable results, despite spectral variations attributable to the hemoglobin absorption peak in the human frozen and unfrozen tissues. However, because the spectral fingerprints of the freeze-thaw process in the ex vivo and in vivo experiments were similar, we could extrapolate the maximum depth of freezing. Therefore, this sensor has the potential to be utilized for monitoring cryosurgery in real time.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.