Abstract

We present a new method for traceable calibration of size and form error of microspheres, which was realised by stitching a series of atomic force microscopic (AFM) images measured at different orientations of microspheres using the metrological large range AFM of the PTB. The stitching algorithm is achieved using an iterative closest point point-to-plane algorithm. As the AFM tip geometry is one of the most significant error sources for the developed method, it was traceably calibrated to a line width standard (type IVPS100-PTB), whose feature geometry was calibrated with a traceable route to the lattice constant of crystal silicon. Measurement setup, scan strategy, and data evaluation processes have been detailed in the paper. Measurement results show high stability and robustness of the developed method. For instance, the standard deviation of four repeated measurements reaches 5 nm, indicating promising performance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.