Abstract

Although many efforts have been put forth, fecal incontinence of colostomates is still an unsolved problem. Among the techniques available to improve continence, the use of electromagnetic/mechanical power has not been explored. Our work looks at the characteristics needed by the components of an artificial sphincter where the power of sphincteric muscles is substituted with electromagnetic/mechanical power, and the mechanical absence of muscles is replaced by a plastic band. Experimental work was done with isolated colon of 26 mongrel dogs, and all of them became continent by an intraluminal gas pressure over 80 mm Hg when a final perimeter between 3.7 and 4.7 cm was reached. No macroscopic lesions were noted on the guts' surface beneath the plastic band. The largest band displacement was 4.2 cm, and the traction force was under 50 N.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.