Abstract

A high-resolution spectrometer has been developed and used to measure simultaneously impurity ion temperature and flow velocities in high temperature plasmas with 10 {mu}s temporal resolution (limited by digitization rate). This device is actually a duo-spectrometer: measurements from two different chordal views of the plasma can be made simultaneously via two separate quartz input fiber optic bundles coupled to the entrance slits which are tilted to compensate for line curvature. The dispersed spectra on the exit plane of the duo-spectrometer are coupled via quartz fiber optics to two arrays of 16 photomultiplier tubes each. Measurement made by recording the Doppler broadened and shifted 227.091 nm emission from the CV impurity ions in the MST reversed-field pinch (RFP) plasma have achieved precisions of <6 eV for temperatures of 150 Ev and <0.7 km/s for flow velocities of 6 km/s. Representative results from the MsT RFP indicate that the toroidal flow velocity drops and ion temperature increases during saw tooth events in MST.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.