Abstract

Personal care product chemicals (PCPCs) are the chemicals used in personal care products. Many of them are endocrine disruptors and have potential adverse effects on humans. The concentrations of PCPCs in urine are the main biomarker for assessing human exposure. A method was developed for the simultaneous determination of 14 PCPCs in human urine using dispersive liquid-liquid extraction combined with ultra high-performance liquid chromatography-tandem mass spectrometry (UHPLC-MS/MS). Compared with liquid-liquid extraction, this method had the advantages of time efficiency, sensitivity, and limited organic solvent consumption. It produced good linearity (0.9965-0.9996), limits of detection (2.82-36.36 pg mL-1 ), limits of quantitation (9.39-121.08 pg mL-1 ), matrix effect (-0.90%-2.55%), intra-day precision (relative standard deviations [RSDs] <15%), and inter-day precision (RSDs <19.9%). The method had satisfactory relative recovery at three concentration levels. A rapid method was developed for the simultaneous quantification of 14 PCPCs in human urine. The practicability of the method was verified with 21 urine from university students. It is expected that this method will provide a powerful reference for the assessment of exposure to PCPCs in large populations.