Abstract

Minimum variance beamformer (MVB) increases the resolution and contrast of medical ultrasound imaging compared with nonadaptive beamformers. These advantages come at the expense of high computational complexity that prevents this adaptive beamformer to be applied in a real-time imaging system. A new beamspace (BS) based on discrete cosine transform is proposed in which the medical ultrasound signals can be represented with less dimensions compared with the standard BS. This is because of symmetric beampattern of the beams in the proposed BS compared with the asymmetric ones in the standard BS. This lets us decrease the dimensions of data to two, so a high complex algorithm, such as the MVB, can be applied faster in this BS. The results indicated that by keeping only two beams, the MVB in the proposed BS provides very similar resolution and also better contrast compared with the standard MVB (SMVB) with only 0.44% of needed flops. Also, this beamformer is more robust against sound speed estimation errors than the SMVB.