Abstract

Sequential cone-beam tomography is a method that uses data of two or more parallel circular trajectories of a cone-beam scanner to reconstruct the object function. We propose a condition for the data acquisition that ensures that all object points between two successive circles are irradiated over an angular span of the x-ray source position of exactly 360 degrees in total as seen along the rotation axis. A fast and efficient approximative reconstruction method for the proposed acquisition is presented which uses data from exactly 360 degrees for every object point. It is based on the Tent-FDK method which was recently developed for single circular cone-beam CT. The measurement geometry does not provide sufficient data for exact reconstruction but it is shown that the proposed reconstruction method provides satisfying image quality for small cone angles.