Abstract
The alignment of a grating-based X-ray phase-contrast interferometer is an iterative process that requires numerous steps of changing the distances and angles between the gratings. For each alignment step an image is acquired to evaluate the detected intensity signature in order to optimize the observed moir'e pattern. Thus, a large number of images has to be taken for the alignment procedure. This is not feasible within reasonable time at X-ray sources like X-ray backlighters where the time between two X-ray shots is on the scale of hours. Here, we report on the development of a stable and transportable setup ready-to-use for grating-based X-ray phase-contrast imaging. A comprehensive set of reference images taken at a continuous beam serves as a look-up table which enables the grating alignment within very few alignment steps. Since this method features a fast, reliable and predictable alignment, it is also beneficial for grating-based X-ray phase-contrast imaging systems at common X-ray sources.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.