Abstract

We present an agile, hardware independent analysis methodology based on the spectral windows technique to detect and quantify the environmental activity concentration of artificial isotopes of interest. The method removes counts from Compton scattering, peak overlapping, intrinsic background and other unconsidered contributions to the spectral region where photopeaks from artificial elements are expected. Taking into account the varying concentration of 214Bi, 214Pb, and 212Pb, we compute the activity concentration of the artificial isotopes (when detectable), or a variable Minimum Detectable Activity Concentration (MDAC). This allows to set statistically robust early-warning levels. In the scenario with the lowest concentration of natural occurring isotopes, and for 10-min integration time, we achieve a MDAC of 5.4/4.1/3.9 (Bq/m³) for 131I/137Cs/60Co using a LaBr3(Ce) detector; and 4.2/2.8/2.7 (Bq/m³) for 131I/137Cs/60Co with a SrI2(Eu) detector.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.